CDF – LEAK/FLOW CALIBRATOR

The Leak/Flow Calibrator allows the checking of the calibration of leak and flow measurement instruments as well as calibrated leaks and jets.

This portable, compact and user-friendly instrument lets you adjust very accurately your leak or flow rate with a real time reading on the display. You can now guarantee the quality of your products, whilst also complying with the requirements of standards such as ISO 9000.

CONTACT US

CDF: APPLICATIONS

• Calibration of ATEQ leak detectors.
• Checking of any other type leak detectors.
• Capability studies and regular calibration check programs.
• Allows checking of leak test fixtures.

CDF: FEATURES

• Measurement in actual or ambient conditions
• USA and S.I. units
• Selection of languages
• Storage of 10 measured values
• Ready for future developments
• Calculation of the mean value
• Possibility to program checking sequence
• Time/Date function
• Automatic sleep mode
• LCD display (4 x 20 characters)
• 1 year’s warranty
• Easy connection to the test instrument
• Can be used in any position
• Precision needle valve
• Robust carry case
• Portable (battery + charger)

MEASUREMENT RANGES

MeasurementRangeAccuracyMaximum Resolution
Leak 1*0 to 2 ml/min± (2.5% of F + 0.003 ml/min)0.001 ml/min
Leak 2*2 to 20 ml/min± (2.5% of F + 0.03 ml/min)0.01 ml/min
Leak 3*20 to 200 ml/min± (2.5% of F + 0.3 ml/min)0.1 ml/min
Leak 4*200 to 2000 ml/min± (2.5% of F + 3 ml/min)1 ml/min
Leak 5* (option)2000 to 20 000 ml/min± (2.5% of F + 30 ml/min)10 ml/min
Leak 6* (option)400 to 4000 l/h± (2.5% of F + 6 l/h)2 l/h
Leak 7* (option)1000 to 10 000 l/h± (2.5 % of F + 15 hPa)5 l/h
PressureEmpty to 10 bar± (1.5 % of the pressure + 15 hPa)0.003 bar
ATM Pressure950 to 1050  mBar± (1 % of X + 3 digits)1 mbar
Temperature+ 15 to 35⁰1 %0.1 ⁰C

 *Only one option at a time for range [4] (choice of: leak 5 or leak 6 or leak 7)
Note: the above specifications are available for a CDF instrument posed on a horizontal surface and inclined with 45° with its rotary handle. F = Flow

CDF: SPECIFICATIONS

  • PHYSICAL

    • Weight : around 3Kg
    • Size : H x W x D: 345 x 192 x 85 mm
  • PNEUMATIC CONNECTIONS

    • STAUBLI type quick release connectors
  • AIR SUPPLY

    • Dry and clean air is compulsory. Air quality standard required ISO8573- 1 class 4.4.3.
  • TEMPERATURE

    • Operating : + 15°C at + 35°C
  • INTERFACE

    • RS232 link via 9 pin subD connector