- Integrated ionizing high voltage generator (5 to 27.3 kV)
- Reject levels as % of nominal voltage (0 to 100%)
- Monitoring and protection of high voltage generator
- Speed: Minimum cycle time 0.6s
- Limitation of the current rating
- I/O’s for instrument control and results
- 32 programs
- Remote control allows the test module to be closed to the test part
- And: Language selection, customization of test …
SPECIFICATIONS
INTERFACE
Programming via remote control 7 inputs / 5 outputs for PLC controlled applications.
- Inputs: Optically isolated.
24 V – 10 mA maximum or dry contact.
- Outputs : Relay output
Rated 48 V / 200 mA maximum.
- Inputs: Optically isolated.
DIMENSIONS
- Box dimensions:
Dimensions: H x L x P = 136 x 250 x 255 mm
Weight: 8 kg
- Remote control:
Dimensions: H x L x P = 250 x 250 x 60 mm
Weight: 2.8 kg
- Box dimensions:
- ELECTRICAL SUPPLY
- 24VDC/1A
- The instrument needs a good ground connection
- TEMPERATURE
- Operating: + 10°C at + 45°C
- Storage: 0°C at + 60°C
OPTIONS
ADD-ONS
- Remote Control
- Save results Module
- Standard resistor box with 2 values
MEASUREMENT PRINCIPLE
- The IONIQ measures the current flowing between a patented charged probe and a ground plate placed under the part to be tested.
- The IONIQ uses the % of the nominal voltage (which reflects the discharge current), measured on the part as PASS/FAIL level.
- In a PASS situation (fig 1) : No hole, no weak part, the IONIQ measures a high %. The measured voltage and the nominal voltage are virtually equal. The result is above the reject level, the part has passed the test.
- In a FAIL situation (fig 2) : The IONIQ measures a low %. The measured voltage is significantly below the nominal voltage. The result is below the reject level, the part has failed the test.
- Test limitations: short probe-part-plate distance, electrical insulation from environment.