APPLICATION
EMOBILITY
AUTOMOTIVE
- Air Conditioning
- Radiators
- Exhaust
- Transmission…
PACKAGINGS
- Aerosol Caps
- Plastic Bags
- Sprays
- Plastic Bottles...
ELECTRONICS
- Hard Disk
- Watches
- TV Screen
- Camera Lenses…
MEDICAL
FEATURES
GENERAL
- Pressure decay leak testing from 5 to 1000 kPa
- Range of measurement from 20 to 175 bar
- 128 programs
- Setting on line (via USB)
- Electronic regulation 4 MPa
- Mechanical regulation 2 MPa/ 5 MPa/ 10.0MPa/ 17.5 MPa
TEST MODES
- Flow units (sccm, mm³/s, cm³/s, cm³/min, cm³/h…)
- Desensitized Test and others, depending on your applications
ENVIRONMENT
- ROHS Standard
FLEXIBLE COMMUNICATION FORMAT
- RS232: Printer, Modbus RTU
- Slave USB: USB key ti transfer data (parameters, results…)
- Master USB: USB key to transfer (parameters, results, …)
- 6 Inputs / 6 Outputs 24 V Programmable for External Automation
PRESSURE DROP MEASUREMENT |
---|
Pressure Range | Measurement Range | Accuracy up to* |
---|---|---|
5MPa Max (50 bar) | 0.01 – 5.00 kPa | +/- 1.0% Rdg + 1%FS |
10 MPa Max (100 bar) | 0.01 – 10.00 kPa | +/- 1.0% Rdg + 1%FS |
From 2 to 20 MPa (20 to 200 bar) |
0.01 – 50.00 kPa | +/- 1.0% Rdg + 1%FS |
From 2 to 20 MPa (20 to 200 bar) |
0.01 > 100.00 kPa | +/- 1.0% Rdg + 1%FS |
From 2 to 20 MPa (20 to 200 bar) |
0.1 – 500.0 kPa | +/- 1.0% Rdg + 1%FS |
From 2 to 20 MPa (20 to 200 bar) |
0.1 – 1000.0 kPa | +/- 1.0% Rdg + 1%FS |
Accuracy: Linearity + Repeatability + Hysteresis *Optional Laboratory Accuracy. Range from 10-100% of Full Scale
SPECIFICATIONS
- PHYSICAL
- Weight: 7 Kg
- Dimensions: (WxHxD) 250x150x360 mm
- ELECTRICAL SUPPLY
- 24VDC/2A or 100-240VAC/50W
- TEMPERATURE
- Operating: + 5°C at + 45°C
- Storage: 0°C at + 60°C
- INTERFACE
- Graphic color screen with simple and intuitive menu-driven operator controls
- SMART KEY: programmable key with user’s functions
- AIR SUPPLY
- Command 4/6 bar
- Pressure 1 or 2 bar > test pressure
- Clean and dry air
- Air quality standard to be applied (ISO 8573-1)
- Command 4/6 bar
TECHNOLOGY
DOCUMENTATION